| Overall performances |
Model |
Mini Autochemistry Analyzer |
| Throughput |
200 tests/h |
| Analysis Method |
End-point, Fixed-time, Rate (Kinetic) |
| Certificates |
CE, FDA, ISO9001, ISO14001, ISO13485 |
Sample & Reagent Systems |
Sample Positions |
37 sample positions |
| Reagent Positions |
28 reagent positions |
| Sample Volume |
2 µl ÷ 70 µl |
| Reagent Volume |
20÷350 µl |
| Probe |
Liquid level sensor function, anti-collision function |
| Probe Washing |
Automatic washing interior and exterior |
| Reagent Cooling |
Refrigerated tray with independent switch |
Reaction System |
Temperature Control |
37+0.1℃, real-time monitoring |
| Cuvettes |
48 reusable cuvettes, optical length 6 mm |
| Mixer Probe |
Independent stirring |
| Washing |
Automatic cuvettes washing |
| STAT function |
YES |
Optical System |
Light Source |
6 V/10 W halogen lamp |
| Spectrophotometry |
Post-Spectral spectrophotometry |
| Wavelength |
340, 405, 450, 510, 546, 578, 630, 700 nm |
| Absorbance |
0÷3,0 Abs |
Calibration & QC |
Calibration |
Linear:K factor, 1-point , 2-point and multipoint liner Non-Linear: Spline, Polygon,Index, Ogarithm,Logit-4P, Logit-5P |
| Quality Control |
Real-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot(2D) |
Data Management |
Software |
Windows 7/8/10 , 32 or 64bit |
| LIS System |
Bi-direction, support HL7 protocol |
| Interface |
LAN port access |
Working Conditions |
Power Supply |
AC220V+10%, 50/60Hz, 110V+10%, 60Hz, 300W |
| Temperature |
10°C÷30°C |
Water Consumption |
3 l/h |
| Humidity |
30÷80% U.R: |
| Dimensional characteristics |
Dimensions (W×D×H) |
625×425×460 mm |
| Net weight |
36 kg |
| Packaging dimensions |
721×531×855 mm |
| Gross Weight |
50 kg |